TROUBLESHOOTING
THE DEVICE DOES NOT TURN ON
lf the device does not turn on, please check (A) whether the power plug has been correctly plugged into the socket and/or whether the device is
turned on. lf the unit is still not turning on, unplug the power plug (B) and unscrew the fuse holder with the inscription "FUSE" on the bottom of the
unit (see picture on the left) completely (with a screwdriver or small coin) and check that the fuse inside is intact. (see picture on the right)..
The fuse is located on the
Intact fuse: the wire in the fuse is
bottom of the device
undamaged.
(C) lf the fuse is intact, screw the fuse holder with the fuse back into the device completely. lf the fuse is damaged, replace the fuse with a
commercially-available fine fuse (TR); 3.15 A; 5 x 20 mm). You are also welcome to contact our customer service. (D) Note that the device may only
be powered by 220V power.
THE DEVICE IS PUMPING, BUT IT ONLY BLOWS ONE OR NONE OF THE TWO CUFFS
lf only one or neither of the two cuffs should inflate after starting the appli cation, please check (A) whether the multi-plug is fully and correctly
inserted into the device. Also, make sure (B) that the air hoses are not kinked, for example, because you are sitting on the hoses. Check (C)
that the gray cuff plugs are all correctly attached to the cuff.
help.uenenengel.de