Maintenance – Replacement of the PURION UV lamp
-
basically, the lamp should be replaced after a service life of about 10.000 hours.
The replacement should be carried out by duly trained personnel.
-
procedure:
Disconnect the UV system from the power supply at all poles using the main switch.
-
Removal of PURION UV-lamp [12]:
•
use a screwdriver for loosening PE-made protective cap's [17] safety screw
•
loosen cable screwing
•
withdraw polyethylene-made protecting cap
•
pull UV lamp by its cable
•
detach the lamp from 4-contact plug (caution: may be hot!)
•
in doing so, touch UV radiator at the socket only (glass-breaking danger!)
EN
-
Cleaning : wipe off any existing fingerprints on the lamp's tube with an alcohol-
soaked rag. Depositions on the quartz tube's inner side may be removed with the
PURION SERVICE KIT entfernen.
-
Reinstalling of the PURION UV-lamp [12]:
•
introduce UV radiator to the 4-contact cut-in unit's cable holder on lampside
(remove fingerprints with an alcohol-soaked rag)
•
introduce UV radiator through the stainless steel head into the quartz-made
cladding tube until the bottom has been reached
•
push transparent protecting cap up to the immersion tube's head sealing
•
fasten lamp's cable by means of cable screwing and implement traction relief
•
push transparent polyethylene-made protecting cap into the sealing of the
immersion tube's head as far as it will go
•
adjust safety screw on the polyethylene-made protective cap [17] by using a
screwdriver
-
Restart the equipment
Hint: Independent of lamps' aging, the UV radiator's socket will turn into a
brownish color under influence of UV light. This will not affect its correct operation.
Liability and warranty
-
For warranty claims, we shall only be liable under the national legal warranty
-
The general terms and conditions of PURION® GmbH are holding
-
We explicitly point out that the warranty does not apply to damage caused by:
• Operating failure in consequence of insufficient following this user information
•
Operating with spare parts such as Lamps and ballasts which are no PURION®
original parts.
•
installation of incompliant accessories
•
incorrect operation / installation
•
removing, manipulating or non-insertion of protective devices
•
improper execution of maintenance
•
wear and non- exchanging of wear agents