12. Advanced application

a) Percentage
When the calibrator is in output mode, short press
accordingly, the 0% or 100% value of each output functionality is as follows:
Output Functionality
Millivolt 100 mV
Millivolt 1000 mV
Voltage
Current
Frequency 200 Hz
Frequency 2000 Hz
Frequency 20 kHz
The 0% or 100% value of each output can be reset by the following method:
1. Press the arrow keys to adjust the value and long press
set as output value.
2. Long press
until the buzzer beeps, a new 0% value will be set as output value.
Note: The 100% value must not be smaller than the 0% value.
Short press
, the output value will add 25% of the range between 100% value and 0% value.
Short press
, the output value will decrease 25% range between 100% value and 0% value.
Note: If you short press
greater than the 100% value and not be smaller than 0% value.
b) Slope
The automatic output functionality of the slope can constantly provide a dynamic signal to the transmitter.
If pressing
, the calibrator will produce a constant and repeated slope (0%-100%-0%). There are 3 kinds of

slope:

1) 0%-100%-0% 40 seconds, smooth slope.
2)
0%-100%-0% 15 seconds, smooth slope.
3)
0%-100%-0% 25% progress slope, each step keeps for 5 seconds.
To exit the slope functionality, press any key except for the slope key.
,
0% value
0 mV
0 mV
0 V
4 mA
0 Hz
200 Hz
2000 Hz
or
to adjust the value of output functionality, the output value shall not be
,
,
to rapidly output percentage value
100% value
100 mV
1000 mV
10 V
20 mA
200 Hz
2000 Hz
20000 kHz
until the buzzer beeps, a new 100% value will be