BDS 2197 Bedienungsanleitung Seite 17

Lackschichten-messgerät für eisen- und nicht-eisen-metalle
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Important note for calibration (deleting the calibration data)
If the calibration fails, first perform the following steps.
Then re-perform all the steps from 2-point calibration.
Push and hold "Zero" button.
1.
Pull trigger
Display shows „Clr" und "Set".
2.
Calibration data be erase.
3.
Start a new 2 point calibration.
Frequent Calibrating Point
Push and hold "MAX/MIN" button.
1.
Pull trigger
Display shows "SET" und "dFut ".
2.
Push „▲" or „▼" button to set calibration value.
3.
example: 39,6 mils (1006 µm)
"CAL" Taste drücken
Calibration value is stored.
4.
Put standard coating plate on measuring plate.
5.
(Standard coating plate 39,6 mils / 1006 micron)
Push sensor on standard coating plate and measuring
plate. Pull trigger.
6.
(Take the ferr or non-ferr measuring plate for calibrate)
Wait until the display shows the
7.
Push „CAL" button
Push "Zero" button for 2 seconds.
8.
Calibration value will be set.
SW-Stahl und Werkzeugvertriebs GmbH
L
e
v
e
k r
u
s
r e
t S
a r
s
s
e
6
5
D
4 -
2
8
9
7
R
e
m
s
c
h
e
d i
icon.
Tel. +49 (0) 2191 / 46438-0
F
ax +49 (0) 2191 / 46438-40
- E
M
a
: l i
n i
o f
@
s
w
s
a t
l h
d .
µm
  
µm
  
µm
  
e

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für BDS 2197

Inhaltsverzeichnis