20 | Instructions for Use
BIF8plus
Different device addresses
If several beds are used with a
in a room, it must be ensured
BIF8plus
3
that each bed only reacts to the associated
remote control. For this
reason, the receiving address can be set by a trained technician. One
of four possible device addresses can be assigned to each
, so
BIF8plus
that up to four
-controlled beds can be in one room.
BIF8plus
Monitoring switch contacts
recognizes when one of the switching contacts is no longer
BIF8plus
electrically separated due to a defect. In this case, the activation of the
switching contacts is deactivated. This faulty condition is indicated by
the
2
red signal light on the front of the case.
System self-monitoring
Safety precautions have been taken in the
to prevent the con-
BIF8plus
trol unit from malfunctioning. In the event of a failure, a watchdog mo-
dule deactivates the activation of the switching contacts.
High protection class
The electronics are protected against water with a housing of protection
class IP65. The power supply is supplied via a short-circuit proof power
supply according to safety standard IEC 60601-1.