Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Specifications - Laica KD-7920 Anleitungen Und Garantie

Handgelenk-blutdruckmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 44
EN
English
Values below 105 mmHg (systolic) and 60 mmHg (diastolic) indicate hypotension. We recommend seeking medical
advice.This device is able to detect irregular heartbeats, or arrhythmia, indicating them on the display with the symbol
. Arrhythmia may be caused by frequent anxiety, certain emotional states, excessive alcohol use, genetic
predisposition, age or other.
It can be a symptom of a particular physical or mental condition (temporary disturbance) or an actual heart problem.
Always seek medical advice if the monitor displays the irregular heartbeat symbol.
PRODUCT DESCRIPTION (SEE FIG.1)
1)
LCD DISPLAY
2)
"M" key
3)
"O/I" key
4)
Cuff
5)
Battery compartment
6)
Batteries
7)
Case

TECHNICAL SPECIFICATIONS

• Product name:
Blood pressure monitor
• Commercial name:
BM1007
• Classification:
Internally powered, Type BF applied part, IP22, No AP or APG, Continuous operating mode
• Method: oscillometric, automatic air inflation and measurement
• Measurement range: from 0 to 300 mmHg (cuff pressure),
from 60 to 260 mmHg (systolic pressure),
from 40 to 199 mmHg (diastolic pressure),
from 40 to 180 beats/min (heart rate)
• Precision:
systolic, diastolic ±3 mmHg
heart rate ±5% of reading
• Memories: 30 memories per area (4 areas)
• Wrist circumference: between approx. 140 and 195 mm
• Power supply: 2 x 1.5V
AAA alkaline batteries (LR03)
• Battery autonomy: approx. 375 readings
• Environmental conditions for use: from +10°C to +40°C; relative humidity ≤85% RH
• Environmental conditions for storage: from -20°C to +50°C; relative humidity ≤85% RH
• Atmospheric pressure for use and storage: 80kPa-105kPa
26
• The blood pressure monitor has been designed to maintain its performance and safety characteristics for a
minimum of 10,000 readings.
The cuff has been designed to maintain its condition for around 1000 opening/closure cycles.
• The components belonging to the pressure measuring system (accessories included): pump, valve, LCD display,
cuff and sensor.
INSTRUCTIONS FOR USE
BATTERY INSERTION/REPLACEMENT
The blood pressure monitor works with 2 disposable 1.5V AAA alkaline batteries.
At first use and when the display shows the battery symbol
Open the battery compartment by sliding the cover in the direction of the arrow, then insert the batteries according to
the indicated poles and close the cover.
Dispose of worn batteries as described in the "Disposal procedure" paragraph.
DATE/TIME SETTING
1) When the device is switched off, press the "M" and "O/I" keys simultaneously for a few seconds: the 24 or 12 hour
time format will flash on the display. Use the "M" key to adjust the date and confirm using the "O/I" key.
2) Then set the year, month, day, hours and minutes. Adjust the data using the "M" key and confirm using the "O/I"
key. Press the "M" key to speed up the number scrolling.
The display shows the hour, minutes and date.
The date/time will need to be reset whenever the batteries are replaced.
WRAPPING THE CUFF
1) Open the velcro wrist strap.
2) Wrap the cuff around the bare wrist as shown in the "Wrapping the cuff" figure and block it by closing the velcro
strap. The edge of the cuff should be about 1-2 cm from the palm of your hand, which should be facing upward,
resting on a surface. The cuff should fit snugly around the wrist but should not be pulled too tightly; leave enough
room to insert a finger between the cuff and wrist. If the cuff is too tight or too loose, the blood pressure readings
will be inaccurate. Never roll sleeves up or the blood flow will be obstructed and the reading will not be accurate.
Make sure that the supplied cuff (14 – 19.5 cm) corresponds to the wrist circumference.
PROPER READING METHOD
For accurate blood pressure readings, follow these instructions:
1) Sit down, relax and keep still for at least 5 minutes before taking the reading.
English
, insert and/or replace the batteries.
27
EN

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Bm1007

Inhaltsverzeichnis