4. AUTOMATIC SLEEP TIME SETTING

4.1 STANDBY

The scale automatically goes into standby mode when we don't have any weight on the platform and after three seconds of being unloaded.
This function is not configurable, it is a fixed function of the normal operation of the indicator.

4.2 AUTO SLEEP

The scale goes into sleep mode to save battery when the scale is idle for a ten minutes or as set in this function. After entering the sleep mode,
any weight placed on the scale will not wake up the indicator. The scale can only be woken up by pressing any keys or plug to the power
source.
After this it will reboot.
This function will act differently if we are only powered by the battery, we will enter the suspension mode in the same way, but if we press
on the platform, we will go from the suspension mode to the standby mode. To return to normal weight mode we will have to press any key as
when connected to the main power supply.
In the weighing mode, press
into the time setting mode. Select a value between 0-99min using the same method as described in Alarm mode digit setting. Press
confirm the value. Now FA SLP will be be displayed again. Press

5. CALIBRATION

ACCESS TO THE TECHNICAL MENU:
Turn on the scale and the selftest will start, before it change into the weighing situation, press
Set the password P 2222,and press
CHANGE THE DIVISION:
Press the key
press
for confirm.
CHANGE THE DECIMAL LOCATION:
Press the key
then press
for confirm.
CHANGE THE MAX CAPACITY:
Press the key
then press
for confirm.
CALIBRATE
to select the F4-CAL , then press
Press the key
ty value can be changed ) and this number keeps flashing, put on 50kg weight, press
unload the weight from the platform, display shows 0.0000, power off the scale, then turn on the scale again, if display shows 0.000, the
calibration is finished
14
and
at the same time, then by pressing
,it will show F1 DIV.
until display shows F1-DIV, Press
until display shows F2-pnt ,Press
until display shows F3-CAP , Press
repeatedly, select FA SLP and press
to get back to normal mode.
to go into the menu, and press the
to go into the menu, and press the
to go into the menu, and press the
to show CAL-1. Then press
OPTIMA USERS GUIDE
three times. Until display shows P 0000.
key to change the division value, then
key to change the decimal location,
key to change the max capacity,
,when display shows50.000 (the max capaci-
key make it stop flashing, display shows 50.000,