Associate SmartGuard with Patient ID
For the first patient use, the SmartGuard requires a one-time association
of the Patient ID. For subsequent measurements, the Patient ID will be
permanently saved on the SmartGuard, which can store up to 168 bilateral
pupil measurements of the associated patient.
11.
There are two options for associating the Patient ID to the SmartGuard.
Select Scan Code to scan the patient's wristband barcode using the
NPi-300 Incorporated Barcode Scanner, or select Manual ID to manually
enter the Patient ID with either alpha or numeric characters.
Scan Barcode Using Incorporated Barcode Scanner
12.
Select Scan Code. The NPi-300 will emit a white light from the top of the
device. Center the light over the barcode until you hear an audible beep.
13.
The Patient ID will now appear on the NPi-300 touchscreen. Confirm the
patient information is correct and select Accept.
Manual Entry of the Patient ID
14.
Select Manual ID. Using the touchscreen or keypad, enter the
alpha or numeric Patient ID and select
information on the screen is correct and select Accept.
Measure Pupils
15.
Position the NPi-300 with SmartGuard at a right
angle to the patient's axis of vision, minimizing
any tilting of the device.
16-17.
Press and hold either the RIGHT or LEFT
button until the pupil is centered on the
touchscreen and the display shows a green circle
around the pupil. A green frame around the
screen indicates the pupil is properly targeted
(Ex. 16), while a red frame indicates the pupil
needs to be re-centered on the screen before
the measurement is initiated (Ex. 17). Once the
green frame appears, release the button, holding
the NPi-300 in place for approximately three
seconds until the results screen is displayed.
18.
Repeat the scan procedure for the patient's
other eye to complete the bilateral pupil exam.
19.
When the bilateral pupil exam is complete, the
NPi-300 measurement results will be displayed
in green for the Right eye and in yellow for
the Left eye. The NPi-300 is defaulted to
open "Results Page 1," displaying NPi and Size
measurements, following the completion of
the Bilateral Pupil Measurement.
20-21.
Using the touchscreen or keypad, select
to view "Results Page 2" with additional
Pupillary Measurement Parameters (Ex. 20).
Select
to view the Pupillary Light Reflex
Waveform (Ex. 21). To return to "Results Page 1"
with NPi and Size, simply select
22.
From the Results screen, select the Video icon
to view the playback of the reading.
. Confirm the patient
15
16
18
20
.
11
12
14
17
19
21