H i k o n e b | 9 0 8 D C
Us e r Ma n u a l
3. Clip and fix the bottle holder (8) to the stand (6) with its screw.
4. If entraining oxygen, connect the oxygen adapter to device outlet, then connect the
bacteria filter (23) to the oxygen adapter (21) and connect the incoming oxygen to O2
adapter. Oxygen adapter must always be connected before the bacterial filter (between
device outlet and bacterial filter). If connected between bacterial filter and ventilation
tube, oxygen will enter to the system without any bacterial filtration.
5. If heated nebulization is desired, then place the heater (19) on chamber between the
lid and the aerosol tubing (15) and connect its power plug to the socket on the back of
the device.
BACK SIDE OF THE DEVICE
Power Input 31
Heater Socket 30
Air Filter 33
Main Fuse 32
9
H N 9 0 8 D C _ E N Re v. A