Herunterladen Inhalt Inhalt Diese Seite drucken

AB CHORUS CI-5292 Bedienungsanleitung Seite 16

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 53
Immunity/
Compliance
Standard
Proximity
Fields from
RF Wireless
Commu-
nications
Equipment
IEC 61000-
4-3
Radiate RF
IEC 61000-
4-3
*Essential performance of the Chorus Sound Processor per
IEC 60601 is defined as auditory stimulation within safe
amplitudes.
RF interference may occur in the vicinity of equipment marked
with the
The Chorus Sound Processor is intended for use in electro-
magnetic evironments with controlled RF disturbances. Elec-
tromagnetic interference may be prevented by maintaining
distance between portable or mobile RF transmitters and the
Chorus Sound Processor, as recommended.
Compliance
level
a
9 - 28
V/m
385
MHz
- 2.45
GHz at
3m
10 V/m
80 MHz
- 2.7
Ghz
80%
AM at 1
kHz
symbol.
Electromagnetic environment - guidance
Proximity RF fields generated
by wireless communications
equipment should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the sound processor, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
15

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis