10. Press the blue Jaw Release button (5) to open the jaws.
11. Push down on the blue Top Lever (1) (closed position) and remove the device from the
cannula.
12. Remove and set aside a spent staple Reload after firing.
13. Reload the device if necessary:
a.
Remove the spent reload by depressing the ribbed grip area on the reload and rotating it
until the alignment pins are disengaged, then pull the spent reload from the handle and
discard it.
b.
Remove the green ship lock from a new JR-REL25-2.0 staple reload by depressing the
ribbed grip area on the reload while rotating the green ship lock in the direction
of the arrow and pulling on it.
c.
Insert a new reload by aligning the blue dot on the reload connector with the blue dot on
the handle's rotation knob and inserting until fully seated.
d.
Rotate the reload shaft clockwise until the blue dot on the outside of the reload shaft
aligns with the blue dot on the rotation knob.
14. The JustRight™ 5mm Stapler may be reloaded and fired up to 15 times in a single procedure.
After Surgery
Warning
Do not reuse or resterilize instruments and accessories labeled "disposable" or "single
use only".
Discard the stapler instrument after use. The instrument is not able to withstand
resterilization. Do not resterilize.