NOTE: Do not remove the green ship lock on the Reload until you are ready to load into the
handle.
3.
Align the blue dot on the reload connector with the blue dot on the handle's rotation knob and
insert until fully seated.
NOTE The reload can only be inserted one way; the tabs on the reload connector are
different widths and match the slots in the rotation knob.
4.
Rotate the reload shaft clockwise until the blue dot on the outside of the reload shaft aligns
with the blue dot on the rotation knob.
Using the Stapler
1.
Insert the instrument into the cannula sleeve.
NOTE: The top blue lever (1) must be in the closed position before attempting to insert the stapler into
cannula sleeve.
2.
Lift the top blue handle (1) to open the jaws.
3.
Turn the rotation knob until the jaws are in the required position.
JustRight Surgical
®
JR-ST25-2.0
Top Lever (blue) – closed
Top Lever (blue) – opened
JR-REL25-2.0
Page 5