Herunterladen Inhalt Inhalt Diese Seite drucken

JustRight Surgical JR-ST25-2.0 Gebrauchsanweisung Seite 6

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
Warning
When positioning the stapler on the application site, ensure that no obstruction, such as
clips, are incorporated in to the instrument jaws. Firing over an obstruction may result in
incomplete cutting action and/or improperly formed staples.
4.
Position the tissue to be transected between the black lines on the outside of the
instrument jaws.
5.
Pull back on the Jaw Clamp Lever (2) until the jaws lock on tissue to be transected.
Warning
If the jaws of the device do not lock closed, the target tissue is outside the compression
specifications. Reposition the jaws on to tissue within compression specification range.
6.
To release and reposition jaws, depress the blue Jaw Release button (5) on the
back of the handle.
7.
Repeat Step 5.
NOTE: After closing, pausing for 10 seconds prior to firing may result in better tissue
compression and better staple formation.
8.
Flip the green Safety Lock (3) down. Do not hold the safety lock down.
Safety Lock (green) - Up
9.
Pull back completely on the blue Staple Deployment and Cut Lever (4) and release to resect
the tissue structure.
NOTE: Only pull back on the blue Staple Deployment and Cut Lever (4) once to completely
fire a staple line.
JustRight Surgical
Black lines
®
JR-ST25-2.0
Safety Lock (green) - Down
JR-REL25-2.0
Page 6