If necessary, clean the nebulizer's atomizers with water; do not use wire wool.
Modifying the size of the atomizers will affect the particle size and the aerosol's therapeutic effectiveness.
Each nebulizer is intended for one patient only.
Aerosol therapy must only take place under medical prescription and in accordance with the prescribed
dosage.
Units are not suitable for use with flammable anaesthetic mixtures with air or with oxygen or nitrous
oxide.
WARNING, for all cases:
Only an approved repair service company is authorised to repair a unit. Always disconnect the power cord
before opening the unit.
Do not modify the unit in any way.
Only use the accessories supplied and listed in the manual.
Do not carry out any servicing or maintenance procedures when the unit is operating.
Contact DTF MEDICAL customer services for all other questions.
Problem
The unit does not make
any operating sounds
The unit makes a lot of
noise when operating
No mist comes out of
the mask or the mouth
piece
Contact DTF MEDICAL or your local supplier:
If one or more parts are missing,
If parts are defective,
For assistance, maintenance or questions about usage,
For problems or unexpected events,
For serious adverse effects.
Warranty
The unit has a 2-year warranty, parts and labour, against any manufacturing defect, except for cases of
improper use. For customer service under warranty, please contact your distributor. Detailed technical
documentation can be provided to you upon request.
7.
Check that the power cord is connected to the mains.
Check that the switch is set to "on".
Check that the current delivered by the power grid matches the current on the unit (Volt
Hz).
Use another device to check whether there is current in the wall socket.
If the problem persists, please contact DTF MEDICAL or your local supplier.
Check that the nebulizer is properly connected to the unit.
If the problem persists, please contact DTF MEDICAL or your local supplier.
Check that the nebulizer contains enough drug product or solution (the minimum amount
for the nebulizer you are using is given in the user instructions).
Check that the tube is properly connected to the unit and the nebulizer.
Check that the tube is not clamped, blocked or twisted.
Check that the atomizer in the nebulizer is properly inserted all the way in.
Check that the atomizer and the air opening in the reservoir are not blocked (clean them
with water).
When using glass vials of drug products, check that no glass fragments have fallen into the
nebulizer.
If the problem persists, please contact DTF MEDICAL or your local supplier.
8.

Warranty / recycling

Troubleshooting

Solution