Reconditioning instruction/Re-use
en
-
The products may be further used.
-
Original spare parts from Etac should always be used.
-
Remember that the service life of the product is limited
-
The products must be cleaned using conventional non-abrasive
cleaning agents (pH value of between 5-9). They must be
cleaned regularly. The products and their accessories can be
autoclaved at 85° for a maximum of 3 minutes and also dried
in a drying cabinet at 70° for 10 minutes (unless otherwise
stated).
1
Check that the leg ferrules are soft, intact and functional. Replace parts if in doubt.
Check that there are no cracks or damage.
2
Check that the seat's leg attachments have not become warped or been tampered with.
Check that the legs stand symmetrically on the floor and have not become warped.
Check that there are no cracks or damage.
3
Check that there are height adjustment buttons on the legs and that they function satisfactorily.
Perform a functionality test of leg height adjustment.
Check that there are no cracks or damage.
4
Make sure that plastic parts have no cracks.
Check that all plastic parts are intact.
General
(see manual). This is particularly important when prescribing
the product to a user who is expected to use it for an extended
period of time.
Washing
D isinfect the product with an alcohol-based disinfectant, 70%.
Ferrules
Legs attachment
Legs height adjustment
Plastic components
4