Herunterladen Inhalt Inhalt Diese Seite drucken

General Specifications; Accessories Supplied; Recommended Accessories - R&S HMC8012 Benutzerhandbuch

Digital multimeter
Vorschau ausblenden Andere Handbücher für HMC8012:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

General Specifications

Power supply:
115 V / 230 V ±10%, adjustable voltage selector 
on rear panel; Frequency 50Hz / 60Hz
Power consumption:
25 W max., 12 W typ.
Ambient temperature:
0 ...+55 °C (operating); –40 ...+70 °C (storage)
EMC compliance:
DIN EN 61326-1, DIN EN 55011
Safety compliance:
DIN EN 61010-1, CAN/CSA-C22.2 No.
61010-1-12
Measuring Category:
CAT II, 600 V; CAT I, 1000 V
Dimensions:
222 x 88 x 280 mm
Weight:
approx. 2,7 kg
Warm-up time:
90 minutes
All data valid at 23°C ±5K after 90 minutes warm-up and with 5 3/4 digits
Additional Specifications
DC voltage
Measuring Method
Sigma Delta analog/digital converter
Input Resistance
>10 GΩ (selectable for ranges 400 mV / 4 V
10 MΩ (on all ranges)
CMRR
120 dB for VCM <500 V, 1 kΩ unbalance in the 
LO-line and 5 measurements/second
SMRR
> 60 dB for 50 or 60 Hz ±0,1% and 5 
measurements/second
Input current
60 pA at 25 °C
Overload protection
1000 V for all ranges
AC voltage
Measuring Method
AC coupled True RMS measurement
Input Resistance
1 MΩ parallel <100 pF (on all ranges)
Crest Faktor
Max. 10 (0,5% additional measurement 
uncertainty)
CMRR
> 60 dB 1  kΩ in the LO-line and frequency 
<60 Hz
Overload protection
750 V
AC current / DC current
Shunt Resistance
13,75 Ω at 20 mA; 1,25 Ω at 200 mA; 25 mΩ bei 
2 A, 10 A
Overload protection
Fuse: F10H250V on the front panel
Resistance
Measuring Method
2-wire and 4-wire
Overload protection
1000 V for all ranges
Continuity
Measuring Method
1 mA constant current
Threshold value
Adjustable in 1Ω steps
Response Time
200 measurements/second
Overload protection
1000 V
Diode Test
Measuring Method
1 mA constant current
Threshold value
Adjustable in 10 mV steps
Response Time
10 measurements/second
Overload protection
1000 V
Temperature
Measuring Method
Resistance measurement with platinum
sensor
Sensor types
PT100, PT500, PT1000
Connection
2-wire, 4-wire
Overload protection
1000 V
Math Functions
Statistics
Min/Max/Average/Standard deviation
Relative measurement
NULL key, offset via keyboard
Logarithmic functions
dB Reference level via keyboard or NULL key 
dBm Reference impedance 50/75/600Ω or 
freely selectable
, 750 V
DC
AC RMS
(for all ranges)
rms
Data logging
Number of measuring 
Internal: 50.000;
counts
external: defined by USB stick capacity. 
Rate Log
Min.: 5 ms typ. (in accordance to measuring
function and resolution); Max: 3.600 s
Duration Log
Internal: 250 s...50.000 h;
External: defined by USB stick capacity
Log Daten
Main, 2nd, Time-Stamp
Interfaces
USB 2.0 (TMC and CDC/VCP), Ethernet
10/100 (LXI in Preparation), IEEE-488/GPIB 
optional
Programming
SCPI, compatible with Agilent 34401A and 
34410A
Reading Rates
Function
Setting
AC Voltage
Slow
Medium
Fast
DC Voltage
Slow
Medium
Fast
AC Current
Slow
Medium
Fast
DC Current
Slow
Medium
Fast
Resistance
Slow
(2-wire)
Medium
Fast
Resistance
Slow
(4-wire)
Medium
Fast
Frequency
Slow
Medium
Fast
Diode
Continuity
Temperature

Accessories supplied:

Line cord, printed operating manual, R&S®HZ15 silicone
test lead with safety connector and test probe, 1m 
(black+red)

Recommended accessories:

R&S®HZ812 
PT100 Temperature probe 2-wire
R&S®HZC95
19" rackmount kit 2RU for R&S®HMC
series
Specifications
Resolution Display
Reading
Rates
(per sec.)
5 3/4
400.000
5
4 3/4
40.000
10
4 3/4
40.000
200
5 3/4
400.000
5
4 3/4
40.000
10
4 3/4
40.000
200
5 3/4
200.000
5
4 3/4
20.000
10
4 3/4
20.000
200
5 3/4
200.000
5
4 3/4
20.000
10
4 3/4
20.000
200
5 3/4
400.000
5
4 3/4
40.000
10
4 3/4
40.000
50
5 3/4
400.000
5
4 3/4
40.000
10
4 3/4
40.000
25
6
999,999
1
5
99,999
10
4
9,999
100
4 3/4
40.000
10
4 3/4
40.000
200
4
999,9
10
63

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hmc8012g

Inhaltsverzeichnis