DIMENSIONS mattress width (VERONA EVO ONLY)
S= 74 cm
2.5. Power supply
The system of power supply must have voltage and frequency equal to the rated voltage and frequency of the power
supply of the device and be appropriate to the installed capacity, as indicated on the identification plate of the table.
Voltage
Frequency
Absorption
Transformer
The power supply must be supplied to the unit by a line:
Protected against overcurrent by an automatic circuit breaker (circuit breaker) or fuses, properly sized; the ope-
rating current of the protections must be as low as possible taking into consideration the maximum absorption
of the device;
Equipped with a device sensitive to insulation failure (differential) with operating current low enough; this device
must be positioned downstream of the overcurrent protective device (or be integrated with it); Having cables
with section suitable for the power absorbed by the appliance.
The table must also be connected to an external unipotential protection circuit (grounding)
adequate.
2.6. Connections
Insert the plug into a suitable power outlet.
If there is incompatibility between the socket and the plug of the table, replace the socket with a suitable type by
professionally qualified personnel.
Pull the plug from the power outlet grasping the plug, not the cable.
6
M= 82 cm
230/110 V
50/60 Hz
0,70 - 1,50
Primary 230V - 110V
Secondary 25V 120VA
XL= 90 cm