Operation
To operate the handpiece, simply depress the foot control.
note: never operate air motor under the following
conditions:
1. at air pressure exceeding 45 p.s.i. (3.1 atm.).
2. On an alternate hose without adjusting air pressure.
automatic Oilers
Since the use of automatic oiling does not eliminate
the need for routine maintenance, the oiler should be
disconnected. Do not use oil mist lubrication.
(Excess oil may hinder performance.)
Speed Range
The Midwest RDH Handpiece offers exceptional torque at speeds
up to 6000 RPM.
Mounting Prophy Right angles
RDH with Right angle attachment only—REF740021)
A plastic head wrench has been provided for mounting or
removing the prophy head.
(for Midwest
10