 Please note that the empty TrayCell has an absorbance of around 1.
To correct this characteristic absorbance, first carry out a reference
measurement with the solvent in which the sample was dissolved.
 You can then start taking measurements with the sample.
 Please note that the cap must be correctly positioned on the TrayCell
before you start measuring!
 If necessary, the sample can be recovered by pipetting once the
measurements are complete.
 Finally, clean the window and the mirror inside the cap.
A
Pipette sample onto window.
C
Remove cap and recover
sample if necessary.
B
Position cap.
Start measurement.
D
Clean window and cap
(TrayCell remains in
spectrophotometer).
Attention: Place the cap in
such a way that the pins are
in the notches of the cap.
E
Pipette new sample
onto window.