Procedure:
- Fill a container, suitable to contain all the individual components to be disinfected, with a solution of drinking water and disinfectant,
while respecting the proportions indicated on the packaging of the disinfectant itself.
- Completely immerse each individual component in the solution, taking care to avoid the formation of air bubbles in contact with the
components. Leave the components immersed for the period of time indicated on the packaging of the disinfectant, and
associated with the concentration chosen to prepare the solution.
- Remove the components now disinfected and rinse thoroughly with lukewarm drinking water.
- Dispose of the solution as recommended by the disinfectant manufacturer.
method B: Disinfect the accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C5-C6 by boiling them in water for 10 minutes; use demineralised
or distilled water to prevent calcium deposits.
method C: Disinfect the accessories C1.1-C1.2.1-C1.2.2-C1.2.3-C2-C5-C6 with a hot steam sterilizer for baby-bottle (not the micro-
wave type). Perform the process faithfully following the instructions of the sterilizer. To ensure that the disinfection is effective, choose a
sterilizer with an operating cycle of at least 6 minutes.
- After disinfection shake the accessories vigorously and place them on a paper towel, or alternatively, dry them with a jet of hot air (e.g.
using a hair dryer).
AIR FILTERING
The unit is equipped with an air filter (A3)
which should be replaced when dirty or when
its colour changes. To replace the filter take it
out by pulling it as shown in figure. The filter
is designed to remain fixed in its housing. Use
original MANUFACTURER accessories only.
ENGLISH
SYMBOLE
Alternating current
CE Medical Marking ref. Dir. 93/42
EEC and subsequent amendments
Important: check the operating
instructions
Class II device
"OFF" for part of equipment
"ON" for part of equipment
Serial number of device
In compliance with European
standard EN 10993-1 Biological
evaluation of medical devices and
93/42/EEC Medical Device Directive.
Phthalates free. In conformity with
Reg. (EC) no. 1907/2006
Type BF applied part
Hazard: electrocution.
Consequence: Death.
Do not use device while taking a
bath or shower
Enclosure protection rating: IP21.
(Protected against solid bodies over
12 mm. Protected against access
with a finger; Protected against
vertically falling water drops.)