Herunterladen Inhalt Inhalt Diese Seite drucken
Vorschau ausblenden Andere Handbücher für MI 1005:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3. Installation
engages with an audible ‚click' (observe installation
direction for MI 2005!). The installation of the
MI 1005/MI 2005 should be done near the medical
appliance/system. The safety advice in Chapter -1.2
– Connection and Operation- is to be observed!

4. Maintenance

For maintaining a safe technical condition, an annu-
al inspection is recommended, which includes the
following checks:
1. Examination for external damage (housing,
network connections, legibility of labeling,
dirt, etc.) of the MI 1005/MI 2005.
2. Examination for penetrated substances,
especially liquids.
3. Function test.
4. Isolation test. The isolation resistance of
the MI 1005/MI 2005 should not exceed
20 MOhm.
For testing, at both connections all data
cables as well as the screen should always
be shut down briefly. The isolation resis-
tance should be measured between signal
entry and signal exit.
5. Availability and sufficiency of the docments.
Page 28
MI 1005/MI 2005 Manual
Rev 3.1 02.2021

Werbung

Kapitel

Inhaltsverzeichnis
loading

Verwandte Produkte für Baaske Medical MI 1005

Diese Anleitung auch für:

Mi 2005

Inhaltsverzeichnis