Herunterladen Inhalt Inhalt Diese Seite drucken

DeFelsko PosiTest AT-M Gebrauchsanweisung Seite 6

Vorschau ausblenden Andere Handbücher für PosiTest AT-M:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4. Zero the instrument BEFORE pumping by pressing the
button. This prepares the instrument for the test by clearing
the display, and zeroing the instrument. The priming icon
will appear (flashing) on the display indicating the instrument
has not achieved priming pressure.
5. Prime the pump slowly until the displayed reading approaches
the priming pressure. The priming pressure is the point that the
instrument begins calculating and displaying the pull rate
(priming icon
the ability to store readings is enabled. Priming pressures for
the various dolly diameters are:
10 mm
14 mm
20 mm
50 mm
50 mm x 50 mm
For optimum results, prior to exceeding the priming pressure,
return the pump handle to its full upright position and then
complete a single continuous stroke at the desired pull rate until
the actuator separates the dolly from the coating.
6. Open the pressure relief valve and remove the dolly from the
actuator assembly.
7. Readings may be stored into memory by pressing the
(memory storage for up to 200 pulls). Press again to review
stored readings. Stored measurements can be accessed using
our PosiSoft Desktop software (pg. 7).
8. To erase all stored test results from memory, press and hold the
button, then press the
from the display
will disappear). It is also the pressure at which
400 psi
200 psi
100 psi
16 psi
12 psi
button. The
5
2.8 MPa
1.4 MPa
0.7 MPa
0.11 MPa
0.08 MPa
button
icon will disappear

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis