Specifications for enclosure port immunity to RF wireless communications
The Product is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Product can help
prevent electromagnetic interference by maintaining radiation levels of RF wireless
communications equipment (emitters) within the compliance limits specified below.
Recommended radiation levels of RF wireless Communications Equipment
Frequency band
380 – 390 MHz
430 – 470 MHz
704 – 787 MHz
800 – 960 MHz
1 700 – 1 990 MHz
2 400 – 2 570 MHz
5 100 – 5 800 MHz
220/222
equipment
IEC 60601-1-2
Test level
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Compliance level
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
Minimum
separation
distance
0.3 m
0.3 m
0.3 m
0.3 m
0.3 m
0.3 m
0.3 m