Step-by-Step Instructions
6.6.5

Completion of the measurements

Before unplugging the measurement cable from the device
receptacle disconnect the lip clip and the connection hook from the
patient.
Move the file stopper to the selected reference point on the tooth.
Gently remove the file from the canal and measure the apical length
between the stopper and the file.
Note
As a safety precaution in order to avoid over-instrumentation, it is
recommended to proceed as follows: stop file progression in the
canal at the point where the Propex Pixi
Place the file onto an endodontic ruler and measure the apical length.
Subtract a minimum of 0.5 mm from the measured file length.
6.7

Sound Adjustment

Propex Pixi
monitoring of the progression of the file within the canal in addition to
visual monitoring.
The volume can be adjusted to four different levels: mute, low, normal
and high, by successively pressing the
When sound level is muted the
levels the
20/222
®
is equipped with an audio indicator which enables
icon remains turned on.
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
®
indicates '0.0'.
"VOLUME" button.
icon is turned off. At other sound