Herunterladen Diese Seite drucken

Beurer medical PO 80 Bedienungsanleitung Seite 3

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

8.1 Function button
The function button on the pulse oximeter has three functions in total:
Switch-on function: When the pulse oximeter is switched o you can hold down the function button to switch it on.
Display function: To select your desired display format (vertical format, horizontal format), press the function button during operation. There are
four di erent display formats to choose from.
Settings menu function: To enter the settings menu, you must fi rst set the display format to a horizontal format. To call up the settings menu,
press and hold down the function button during operation. You can set the following parameters in the settings menu: display brightness, alarm set-
tings, activate/deactivate data transfer, data storage and upload data to a PC.
Note
To be able to set the following functions (sections 8.2 to 8.6), you must select a horizontal format as the display format (see 8.1 Function button /
Display function). If the display format is set to a vertical format and you would like to use the following functions, an error message appears on the
display ("Please change direction fi rst!").
8.2 Display brightness
To set the display brightness, switch on the pulse oximeter and press and hold down the function button. In the settings menu, the menu item
"Brightness" is selected.
There are four di erent brightness levels to choose from. To set the desired display brightness, press and hold down the function button until the
desired level of brightness has been reached.
To exit the settings menu, use the function button to select the "Exit" menu item and confi rm by pressing and holding down the function button.
8.3 Alarm settings
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
In the settings menu, use the function button to select the "Alarm" menu item and confi rm by pressing and holding down the function button.
Use the function button to select the desired parameter and set the desired value by pressing and holding down the function button.
Y
ou can set the following parameters in the alarm menu:
"Dir"
"SPO2 ALM HI"
"SPO2 ALM LO"
"PR ALM HI"
"PR ALM LO"
"Alarm"
"Pulse Sound"
To exit the alarm menu, use the function button to select the "Exit" menu item and confi rm by pressing and holding down the function button.
8.4 Displaying measurement values in real time on a PC ("USB")
Trend
Current
wave form
of the
pulse rate
Measurement posi-
tion number & view
Connect the pulse oximeter to your PC using the supplied USB data cable.
Start the "SpO -Viewer" program on your PC. If you have not yet installed the program, follow the instructions in section 7.2 (Installing the "SpO -
Viewer/Manager" software).
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
Select the "USB" menu item using the function button.
Press and hold down the function button. The word "on" is displayed in the "USB" menu item. Insert one fi nger into the fi nger opening of the pulse
oximeter. After a few seconds, the measurement data is displayed on your PC in real time.
If you disconnect the data cable from the computer or close the program, an entry mask appears on the PC screen. Here you can enter personal
details and store the current measurement data. As standard, the measurement data is stored under C:/Programs/SpO2/Data.
If you no longer want to display the measurement data on your PC, press and hold down the function button. The word "o " is displayed in the
"USB" menu item.
To exit the settings menu, use the function button to select the "Exit" menu item and confi rm by pressing and holding down the function button.
8.5 Recording measurement data
With the pulse oximeter PO 80, you can record your measurement data over a period of up to 24 hours. If required, the measurement data can be
stored on your computer or printed out as a report.
To record the measurement data, follow these steps:
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
Select the "Record" menu item using the function button.
Press and hold down the function button. The word "Time" will appear on the display. Here you can enter the current time of day so that, later on in
the PC software program, you are able to read o the time you started your recording.
Time;
hours
To enter the current time of day, press and hold down the function button and confi rm your entry by pressing the function button briefl y.
Note
If you start a new recording, the previous recording is automatically overwritten and cannot be recalled. The maximum recording duration is 24
hours.
To start the recording, use the function button to select the "Y" on the display and confi rm by pressing and holding down the function button. The
pulse oximeter starts recording. If you do not want to start recording, select "N" and confi rm by pressing and holding down the function button.
Note
If you have started a recording and return to the display, a red recording symbol appears (REC •). To save battery power, the device automatically
switches to energy-saving mode 30 seconds after the recording starts. If the pulse tone is activated, it is automatically deactivated.
If you press the function button briefl y when in energy-saving mode, "Recording" appears on the display for 2 seconds.
If you press and hold down the function button when in energy-saving mode, the normal display appears.
"Memory is full" appears on the display when the pulse oximeter memory is full.
8.6 Uploading measurement data to a computer
Note
You cannot transfer measurement data to the computer while a measurement data recording is underway.
You can upload the measurement data that you previously stored on the pulse oximeter to your computer.
To upload your stored measurement data to your computer, follow these steps:
Connect the pulse oximeter to your PC using the supplied USB data cable.
Start the "SpO -Manager" program on your PC.
Options
Click the magnifying glass symbol in lower section of the program. The "New Session" window opens.
Please enter your user data. To confi rm the user data, click "ok". The program now displays that the pulse oximeter is connected to the computer
and that the program is ready to receive the measurement data ("Device connected : wating for data ...").
Switch on the pulse oximeter.
Press and hold down the function button on the pulse oximeter. The settings menu appears.
Select the "Upload" menu item by pressing the function button briefl y.
Briefl y press the function button. The word "on" is displayed in the "Upload" menu item. Your measurement data are now transferred to the com-
puter. It is not possible to cancel the transfer of the measurement data manually.
When the transfer of the measurement data is complete, the "Exit" menu item is highlighted automatically. To exit the settings menu, press and hold
down the function button.
As soon as your measurement data has been transferred to the computer, your measurement data appears in the software program.
After the measurement data has been transferred successfully to the PC, you can also print out the data in the software program, if required.
To do so, follow these steps:
Click the print symbol in the bottom right-hand corner of the program. The "Report" window opens.
Downloaded from
www.Manualslib.com
Here you can set whether the setting value runs up or down when setting the alarm limits in the alarm menu. It is nec-
essary to change the setting direction if you would like to move the limits up or down.
Here you can set an upper limit for oxygen saturation. If, during a measurement, the set limit is exceeded, the satura-
tion value appears yellow and a signal sounds (if the alarm is activated).
Here you can set a lower limit for oxygen saturation. If, during a measurement, the set limit is undercut, the saturation
value appears yellow and a signal sounds (if the alarm is activated).
Here you can set an upper limit for the pulse rate. If, during a measurement, the set limit is exceeded, the pulse rate
appears yellow and a signal sounds (if the alarm is activated).
Here you can set a lower limit for the pulse rate. If, during a measurement, the set limit is undercut, the pulse rate ap-
pears yellow and a signal sounds (if the alarm is activated).
Here you can activate ("on") or deactivate ("o ") the alarm. If you have activated the alarm and the set upper or lower
limit is exceeded or undercut, a signal sounds.
Here you can activate ("on") or deactivate ("o ") the pulse tone. If you have activated the pulse tone, a signal sounds at
every beat during the measurement.
Note
You can create a printout at any time during measurement
data recording by using the print function. The printout shows
the trend of the current measurement up to that point.
Note
You can view the measurement data of up to four devices at
once in the software program. In the software program, you
can enlarge the view of the individual measurement data by
selecting the corresponding measurement position number.
Trend display time
Report
Time;
Start record-
Do not start
minutes
ing
recording
glass symbole
Report
Trend display time
(New Session)
Note
When printing the measurement data, you can choose between four di erent display
variants (Summary Report, Full Study Report, OxiMetry Report, Strip Chart Report).
Together with the printout, the user is o ered an automatic appraisal of the recorded
measurement data. This printout serves exclusively to support the diagnostic fi ndings
of your attending doctor. The printout is not a substitute for a medical certifi cate and is
not exhaustive!
manuals search engine
8.7 Calling up stored measurement data
To call up previously stored measurement data in the software program, follow these steps:
Start the "SpO -Manager" program on your PC.
Click the magnifying glass symbol in the lower section of the program. Select "Review" in the upper section of the window and click "Find File" in
the lower section.
Select the fi le you stored (as standard, the fi les are located under C:/Programs/SpO2/Data) and click "Open" followed by "Ok".
Your stored measurement data is now displayed in the software program.
9. Evaluating measurement results
Decline in oxygen saturation depending on altitude
WARNING
The following table for evaluating your measurements does NOT
Note
apply to people with certain pre-existing conditions (e.g. asthma,
The following table informs you of the e ects of various altitudes on oxygen
heart failure, respiratory diseases) or whilst staying at altitudes
saturation value and its impact on the human body. The following table does
above 1500 metres. If you have a pre-existing condition, always
NOT apply to people with certain pre-existing conditions (e.g. asthma, heart
consult your doctor to evaluate your measurements.
failure, respiratory diseases etc.). People with pre-existing conditions can show
signs of illness (e.g. hypoxia) at lower altitudes.
SpO (oxygen satura-
Classifi cation/measures to be
tion) measurement
Altitude
taken
in %
99-94
Normal range
1500-2500 m
Decreased range:
94-90
2500-3500 m
visit to the doctor recommended
Critical range:
< 90
3500-5800 m
seek medical attention urgently
5800-7500 m
7500-8850 m
Source: Hackett PH, Roach RC: High-Altitude Medicine. In: Auerbach PS (ed):
Wilderness Medicine, 3rd edition; Mosby, St.Louis, MO 1995; 1-37.
10. Maintenance/cleaning
IMPORTANT:
o not use high-pressure sterilisation on the pulse oximeter!
D
Under no circumstances should you hold the pulse oximeter under water, as this can cause liquid to enter and damage the pulse oximeter.
Clean the housing and the interior rubber surface with a soft cloth dampened with medical alcohol after each use.
If a low battery status appears on the display of the pulse oximeter, change the batteries.
If you are not going to use the pulse oximeter for more than one month, remove both batteries from the device to avoid possible leaking.
11. Storage
IMPORTANT:
Store the pulse oximeter in a dry place (relative humidity ≤95 %). If the humidity is too high it may shorten the service life of the pulse oximeter or dam-
age it. Store the pulse oximeter in a place where the ambient temperature is between -40°C and 60°C.
12. Disposal
Dispose of the device in accordance with Directive 2002/96/EC – WEEE ("Waste Electrical and Electronic Equipment"). If you have any queries,
please contact the appropriate local authorities.
The empty, completely fl at batteries should be disposed of through specially designated collection boxes, recycling points or electronics retailers.
You are legally required to dispose of the batteries.
Note: the codes below are printed on batteries containing harmful substances: Pb = Battery contains lead, Cd = Battery contains cadmium,
Hg = Battery contains mercury.
13. What if there are problems?
Problem
Possible cause
The pulse oximeter is not displaying meas-
The pulse oximeter has run out of battery.
urement values
Insu cient circulation in the measurement
fi nger
The pulse oximeter is displaying measure-
Measurement fi nger is too large or too small
ment interruptions or high measurement
value jumps
Finger, hand or body is moving
Cardiac arrhythmia
14. Technical Data
Model no.
PO 80
Measurement method
Non-invasive measurement of arterial oxygen saturation of haemoglobin and pulse rate in finger
Measurement range
SpO 0 – 100%,
Pulse 0 – 254 beats /minute
Accuracy
SpO 70 – 100%, ± 2%,
Pulse 30-250 bpm, ± 2 beats /minute
Dimensions
L 57 mm x W 32 mm x H 30 mm
Weight
Approx. 42 g
Sensor to measure SpO
Red light (wave length 660 nm); infra-red (wave length 905 nm); silicon receiver diode
Permissible operating conditions
+10 °C to +40 °C, ≤75 % relative humidity, 700 –1060 hPa ambient pressure
Permissible storage conditions
-40 °C to +60 °C, ≤95 % relative humidity, 500 –1060 hPa ambient pressure
Power supply
Integrated, rechargeable lithium battery, 500 mAh / 3.7 V
Classification
IP22, application part, type BF
System requirements for software
Supported operating systems: Windows XP, Windows Vista and Windows 7
Technical information is subject to change without notifi cation to allow for updates.
This device complies with European Standard EN60601-1-2 and is subject to particular precautions with regard to electromagnetic compatibility.
Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Cus-
tomer Service address or found at the end of the instructions for use.
This device complies with the EU Directive 93/42/EC concerning medical devices, the Medizinproduktegesetz (German Medical Devices Act) and
the DIN EN ISO 9919 standard (Medical electrical equipment – Particular requirements for the basic safety and essential performance of pulse
oximeter equipment for medical use)
Guidance and manufacture's declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration –electromagnetic emission
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the PO80 Pulse
Oximeter should assure that it is used in such an environment.
Emission test
compliance
Electromagnetic environment-guidance
RF emissions
Group 1
The PO80 Pulse Oximeter uses RF energy only for its internal func-
CISPR 11
tion. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
Class B
The PO80 Pulse Oximeter is suitable for use in all establishments,
CISPR 11
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
Harmonic emissions
Not applicable
used for domestic purposes.
IEC 61000-3-2
Voltage fl uctuations/
Not applicable
fl icker emissions
IEC 61000-3-3
Guidance and manufacture's declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture's declaration-electromagnetic immunity
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment specifi ed below. The customer or the user of PO80 Pulse
Oximeter should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
±6kV contact
±6kV contact
±8kV air
±6kV air
Power frequency (50 / 60Hz) magnetic fi eld
3 A/m
3 A/m
IEC 61000-4-8
Expected SpO
value (oxygen satu-
Impact on human body
ration) in %
> 90
No altitude sickness (normally)
Altitude sickness, acclimatisa-
~90
tion recommended
Very frequent altitude sickness,
<90
acclimatisation absolutely
essential
Severe hypoxia, only limited
<80
length of stay possible
<70
Immediate, acute danger to life
Solution
Charge the battery via the USB connection.
Observe the warnings and safety notes in
section 5
Fingertip must have the following measure-
ments: Width between 10 and 22 mm
Thickness between 5 and 15 mm
Keep your fi nger, hand and body still during
the measurement.
Seek medical attention
Electromagnetic environ-
ment-guidance
Floors should be wood, con-
crete or ceramic tile. If fl oor are
covered with synthetic material,
the relative humidity should be
at least 30%. The manufacturer
may recommend the ESD pre-
cautionary procedures to user.
Power frequency magnetic fi elds
should be at levels characte-
ristic of a typical location in a
typical commercial or hospital
environment.

Werbung

loading