2.5
Zero window
3.0
Tag Info
3.1
Version
3.2
Key lock
3.3
Reset Data
4.0
Password
Here you can define a range around the zero
point (0 bar) in which the display and output
value (0 bar) is set to zero bar. This can be set
for max. 33% of the factory range. Outside the
entered window, the measured value is
approximated:
Display of the tag number (can only be changed
using the software)
Version status of the firmware
A time lock in the steps off, 10, 20... 60 minutes
for keyboard operation can be preselected
here. If "no" is entered, there is no lockout. The
keyboard is unlocked by switching the device
off and on again.
To cancel all user settings, set this parameter to
"yes". Please note: The parameterization is
reset at the moment you set the parameter to
"yes" and confirm it.
accessible for service only