6

Control Structure

Menu Item
Display
0.0
1.0
Language
1.1
Setting Measured
value display
2.0
Unit
2.1
Measuring range start
2.2
Measuring range end
2.3
Offset
2.4
Damping
Description / Adjusment.
Current Measured Valuet
The following display languages are available:
German, English, French, Italian, Spanish,
Portuguese, Dutch
Pressure / Temperature
Pressure
Contents / temperature (only appears if a table
of contents is stored via software)
Content (only appears if a table of contents is
stored via software)
bar, mbar, mmWs
Enter the measuring pressure for 4mA output
signal here
Enter the measuring pressure for 20mA output
signal here.
Correct the measuring pressure at zero point
here. The current measured value is displayed,
which you can change with the arrow keys by
up to +/-33% of the basic range
To calm down the pressure measurement, set
the minimum rise time in 1/10 sec. here.
(Adjustable 0...20 sec.)