Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Message (Display)
TEST FINISHED. BATTERIES ARE
DEFECT, PLEASE CHANGE
TESTING BATTERIES...
BATTERY HEALTH IS SUFFICIENT
TEST FINISHED.
BATTERY HEALTH IS SUFFICIENT
12.9

Step Counter

The controller saves and displays the steps that your patient has made
with the leg with orthosis.
If both legs of the patient are treated and settings are made
for one leg, the currently selected menu item and the selected
leg are displayed alternately.
The step counter is available to you as an orthotist or qualified/trained
expert via the multi-purpose device as well as the Expert app and to the
patient via the User app.
In order to have the steps displayed in the multi-purpose device, select the menu item
steps made with the orthosis in the different modes (A = Auto mode, F = Free mode, L = Lock mode) will be
displayed (fig. 28). For the total number of steps, double the indicated value.
48
Meaning
The test is finished after 2 hours
BACK
maximum. The battery health
is bad.
The test is finished after 2
BACK
hours. The battery quality is
sufficient.
The test is finished after 5
BACK
hours. The battery quality is
sufficient.
Further Action
Exchange the batteries for new
ones.
You can cancel the test and use the
batteries or wait until the test is
finished.
The test has shortened the period
of use of the batteries. Charge the
batteries completely before using
them again.
STEPS WALKED
WITH ORTHOSIS LEG: x
A: x F: x L: x
BACK
Step Counter. The
fig. 28

Werbung

loading

Verwandte Produkte für FIOR & GENTZ ET3840-T