The following messages can be displayed when the test is started:
Message (Display)
ERROR: PLEASE PUT IN FULLY
CHARGED BATTERIES
TRY AGAIN
TEST WITH TWO SOLENOIDS
IS IMPOSSIBLE. DISCONNECT
ONE SOLENOID.
TRY AGAIN
TESTING BATTERIES...
CANCEL
The following messages can be displayed during or after the test:
Message (Display)
BATTERY TEST INTERRUPTED
BACK
Meaning
CANCEL
The batteries in the controller
are not fully charged.
You have mounted two system
knee joints in the orthosis. The
CANCEL
battery health test cannot be
carried out with two connected
solenoids.
The battery health is being
tested.
Meaning
The test has been interrupted,
e.g. because the batteries have
become loose from the holder or
the battery health is too bad to
continue the test.
Further Action
Please insert fully charged batteries
in the controller and select TRY
AGAIN.
Disconnect one solenoid.
Wait for 2 hours and check the
display again.
Further Action
Exchange the batteries for new ones
and make sure that the controller/
orthosis is stored in a way that the
batteries cannot become loose from
the holder.
Repeat the test.
47