Installation of the assembled filter/opener device
Step 3
I
Puncturing of the capsule
Keep the frontend pointing downward. Gently screw the Freezpen
backend to the point of contact resistance. Now a last turn must be
made in an rapid uninterrupted rotation. You may eventually hear a
short hissing sound. The shorter the hiss, the better!
Step 4
I
Use your Freezpen
Remove the applicator cap. Push the trigger to start the flow of
cryogen. The releasing of the trigger will close the flow of cryogen.
You may observe that after releasing the trigger, the remaining gas
in the applicator is discarded.
The working time of one 8g capsule is
95 seconds of continuous flow at a room
temperature of 19°C.
The working time of one 16g capsule is
190 seconds of continuous flow at a room
temperature of 19°C.
Position of the Freezpen
With respect to the content of the cartridge hold the instrument
vertically with the tip directed downward. An angle of less than 45°
must be avoided.
Effective application
Before starting treatment consider the pathology, the location,
the size, the depth, the skin type, degree of tanning the season of
the year and other parameters that may influence the result of the
treatment.
While using Freezpen the applicator tip with the jet of cryogen
9