Herunterladen Inhalt Inhalt Diese Seite drucken

After Use; Power Supply - Dräger X-act 5000 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-act 5000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Use
3.3

After use

3.3.1
Switching off the Dräger X-act 5000
When the instrument is switched off, it is automatically flushed to
remove any reactive gas, which may have been produced by the
chemical reaction of the tubes.
NOTICE
i
i
The instrument cannot be turned off in the following cases:
during a self test,
during a leak test,
during a measurement,
when the password-protected menu is activated.
1. To switch off the instrument, press the "UP" and "DOWN" keys
simultaneously for 3 seconds until the numbers 3 ... 2 ... 1 are
displayed.
After this, a flush cycle will automatically be started. It will take
approximately 5 - 10 seconds until flushing is completed. A
countdown will be displayed.
Then the instrument will automatically be switched off.
18
3.4

Power supply

WARNING
!
To prevent ignition of flammable or combustible
atmospheres, and to avoid compromising intrinsic safety of
the instrument, read, understand, and adhere to the following
procedures. Only use a power supply detailed in the order list
(see chap. 9 on page 25). Otherwise, the approval will be
invalid and there is the risk of explosion.
3.4.1
Charging the rechargeable battery
WARNING
!
Do not charge the rechargeable battery in areas subject to
possible explosion. Use only charging power supplies
specified by Dräger.
NOTICE
i
i
The Dräger X-act 5000 does not function during charging.
The rechargeable battery can either be charged while inside the
instrument or separately from the instrument.
Connect the charger to both
the battery charger connection
(1) and the mains supply.
2
1
00621900.eps
Dräger X-act 5000

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis