Herunterladen Inhalt Inhalt Diese Seite drucken

Data Hold Probe Pm9263; Technical Data; Pt-100 Temperature Probe Pm9249 - Philips PM2521 Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

27
3.2.2.3. Data Hold probe PM 9263 {Fig. 5page 126)
INTRODUCTION
The PM 9263 is a DATA HOLD probe that can be used in combination with multimeters having data hold
facilities on the DIN probe input. A switch ring on the probe is pushed forward to hold the data for display.
Depending on the multimeter, voltage, resistance and current measurements can be made in combination with
the probs-
»
TECHNICAL DATA
Maximum input voltages:
Data Hold by means of slide switch
Probe tip (VL2) to common (0)
30 V rms
VHz product < 10^
on the probe
Common (0) to earth
30 V rms
Temperature range
Probe tip (Vf2) to earth
30 V rms
Rated range of use —10 °C...+55 °C
V test
+
500 V ac
Limit range of storage
and transport
—25 °C...+70 °C
Maximum input current
200 mA
Relative humidity 10%...80%
Input capacity
300 pF
(non-condensing)
Resistance VL2 and 0 leads
130 M£2
ACCESSORIES
Delivered with the PM 9263
Accessory box containing:
• Zeroing lead . 3
• Insulating cap .
8
• 6 markings rings (red, white, blue). 4
• Dual-in-line cap.
9
• Probeholder. 5
• 10 soldering test points. 10
• Spring loaded test clip. 6
• 2 Space probe tips . 11
• Wrap pin connector . 7
*» • Instruction manual
Optional accessories
Accessories used with oscilloscope probes can be combined with this probe.
3.2.2.4. Pt-100 temperature probe PM 9249 (Fig. 6 page 126)
*
The Pt-100 temperature probe is a contact probe, suitable for measurement of surface temperatures between
-60 °C and +200 °C.
Range
-60 °C to+200 °C
Accuracy (DIN 43760) Probe
- 60 °C to+100 °C ± 0,55 °C
(Excluding accuracy of the PM 2521)
+100 °C to+200 °C ±
1 °C
V