2.1.11. Trigger level
Ranges
V
2
20
200
2000
Number of representation units
2100
Accuracy,
Range 2 V and 20 V
(valigPbetween 5 % and 100 % of range)
Frequency
Accuracy
DC
- 500
Hz
± ( 0,3 % of reading +0,5 % of range)
500
Hz -
20 kHz
+ ( 1
% of reading +0,5 % of range)
20 kHz - 100 kHz
± ( 5
% of reading +0,5 % of range)
100 kHz -
10 MHz
± (30
% of reading +0,5 % of range)
Range 200 V and 2000 V
Frequency
Accuracy
DC
- 100
Hz
100 Hz
-
10 kHz
± ( 0,3 % of reading +0,5 % of range)
+ ( 5
% of reading +0,5 % of range)
Temperature coefficient
Input impedance
Maximum input voltages
Trigger level adjustment
Trigger level polarity
+ 100/uV /°C
Range
Impedance
''
2 V
20 V
200 V, 2000 V
20 MO//60 pF
11 M£2/785 pF
10 M£2/795 pF
In all ranges:
Between Hi and Lo
1000 V rms
Between Hi and earth
1000 V rms
Between Lo and earth
400 V rms
Maximum VHz product
10^
*
With thumbwheel at the front of the PM 2521 between
0 and end of range
+ or — selected with +/— pushbutton at the front of the
PM 2521
*