Herunterladen Inhalt Inhalt Diese Seite drucken

Error Messages - Soehnle Professional 6858 Gebrauchsanweisung

Tierwaage
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

14. ERROR MESSAGES

During the initial power-on testing or during the operation the scale may show an error message.
The error messages are described below.
ERROR
DESCRIPTION
CODE
A continuous beep is
nnnnn
heard.
If the selected mass is
less than 10% or more
than 20% of the capacity
CALEr
of the scale, an error
message "CALEr" will be
displayed and the scale
will return to zero.
If an error message is shown, repeat the procedure that caused the message such as turning the
scale on, calibration or any other functions. If the error message still appears, contact Soehnle
Industrial Solutions customer service.
POSSIBLE CAUSES
Weight on the pan exceeds the capacity of the scale.
Remove the weight from the weighing platform.
Incorrect calibration mass. Repeat the process correctly.
14

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis