Generali Information
Designed and created in accordance with the latest standards, FASTERJET is an
highly efficient in medical therapies administered by means of aerosol nebulization. FASTERJET is equipped with a double-valve
system(on top of the nebulizer and mouthpiece),
Safety Warnings
• A single-patient use is recommended • The device should be used only with compressor nebulizers for aerosoltherapy that comply with
standards in force and whose use is indicated in the instruction manual • Device not suitable for anaesthesia and lung ventilation
• The device should be used only with original accessories shown in this manual • Information concerning connection and use with
devices for aerosoltherapy are shown in the instruction manual of the devices. Always observe instructions and safety warnings as
indicated in the instruction manuals of the devices for aerosoltherapy • Use the device and its accessories as per your doctor's
recommendations • Use medications prescribed by your doctor according to his indications • Do not use the device for other use than
Nebulizer for Aerosoltherapy. Manufacturer will not be held responsible for inappropriate use of the device • The device does not come
in sterile pack; always carry out cleaning and sterilization operations before and after each treatment. When using the device follow
ordinary hygiene and cleaning precautions, (wash your hands and carry out cleaning operations as described in the specific section of
this manual) • The device has small components which might be removed and easily swallowed. Use by minors and disabled people
require presence of an adult with his faculties. Do not leave the device unattended in places easily accessible by minors and disabled
people • Always store the device in a dry and clean place far from light, heat and weather • Dispose of device as provided by standards
in force • In presence of pathologies with microbial contamination and infection hazard a single-patient use of accessories and nebulizer
is recommended (always consult your doctor)
How to prepare and to operate the device
The device is not sterile. Before use carry out cleaning and disinfection operations as described in the specific section
To pour the medication rotate the nebulizer top counter clockwise and remove it. Pour the quantity of medication prescribed by your
doctor into the nebulizer tank. IMPORTANT: In case you do so, empty the nebulizer, clean it thoroughly and proceed with new
filling. When medication has been poured, connect the nebulizer top onto the tank, rotate clockwise and insert the accessory prescribed
by your doctor into the nebulizer top. Firmly connect one end of the air tube to the nebulizer and the other end to the air outlet on the
compressor nebulizer for aerosoltherapy. Start the compressor for treatment to begin.
IMPORTANT: NEVER INHALE IN HORIZONTAL POSITION. NEVER BEND THE NEBULIZER OVER 60°.
Cleaning instruction
After each treatment clean thoroughly each component of the nebulizer removing medication residuals and possible impurities. Clean
each component as described here below except the air tube which -in case of treatment to a new patient or in case of impurities should
be replaced. Wash all components (except air tube) under tap warm water (max 60°C) for about five minutes adding if necessary
a small quantity of detergent following dosage and use limitations as provided by detergent manufacturer. Rinse thoroughly making sure
that all deposits are washed away and let dry. All nebulizer components can be disinfected (except air tube) with chemical disinfectants
following dosage and use limitations as provided by disinfectant manufacturer . They are usually available at pharmacy shops (Milton®,
Amuchina® etc.).All nebulizer components can be heat steam sterilized (except air tube and masks) to 121°C (20 min.) or 134°C
(7 min.) (EN554/ISO11134). The Sterilisation packaging must be conform to EN868/ISO11607 and be suitable for steam sterilisation.
After sterilization always let all nebulizer components cool down to ambient temperature before use. Do not repeat sterilizatio n
cycle when nebulizer components are still warm.
Nebulizer life
It is advisable replace FASTERJET nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization
cycles
Storage and transportation conditions
Storage and transportation temperature:
Storage and transportation humidity percentage:
Operatinh technical data
Operating Minimum flow
Operating pressure
Operating Max flow
Operating pressure
Min. capacity
Max. capacity
Read carefully instructions for use
FASTERJET is a Medical Device in conformity with 93/42/EEC Directive
0434
regulating the flow and to dramatically reduce the waste of medication during therapy.
- 20°C (-4°F) ÷ +70°C (158°F)
5% ÷ 95% RH
5 l/min
60 kPa (0.6 bar, 8.7psi)
7.5 l/min
150 kPa (1.5 bar, 21.75psi)
2 ml
16
ml
14
Medical Device (Risk Class IIa - MDD 93/42/EEC)
+/- 10%
+/- 10%
+/- 10%
+/- 10%