Herunterladen Inhalt Inhalt Diese Seite drucken

Allied EPV200 Bedienungsanleitung Seite 18

Tragbares beatmungsgerät (sauerstoffquelle erforderlich)
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

18
Approximate Settings Based on Patient Height
2 Second Inspiratory Time
Tidal Volume
Height Male
=((IBW-50)/2.3)+60
inch
@10 ml/Kg
Height Female
=((IBW-
45.50)/2.3)+60 inch
@10 ml/Kg
Flow Rate
8
BPM = I:E (1 TO - )
9
BPM = I:E (1 TO - )
10
BPM = I:E (1 TO - )
11
BPM = I:E (1 TO - )
12
BPM = I:E (1 TO - )
13
BPM = I:E (1 TO - )
14
BPM = I:E (1 TO - )
15
BPM = I:E (1 TO - )
16
BPM = I:E (1 TO - )
17
BPM = I:E (1 TO - )
18
BPM = I:E (1 TO - )
19
BPM = I:E (1 TO - )
20
BPM = I:E (1 TO - )
Ideal Body Weight
Kg
@10 ml/Kg
Flow Rates and I:E Ratios
By Tidal Volume and Breaths Per
Minute
400
480
560
56
59
63
4' 8"
4' 11"
5' 3"
58
61
65
4' 10
5' 1"
5' 5"
12
14.4
16.8
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
40.0
48.0
56.0
640
720
800
66
70
73
5' 6"
5' 10"
6' 1"
68
72
75
5' 8"
6' 0"
6' 5"
19.2
21.6
24
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
64.0
72.0
80.0
17
880
960
1040
77
80
83
6' 5 "
6' 8"
78
82
85
6' 8"
6' 10"
26.4
28.8
31.2
2.8
2.8
2.8
2.3
2.3
2.3
2.0
2.0
2.0
1.7
1.7
1.7
1.5
1.5
1.5
1.3
1.3
1.3
1.1
1.1
1.1
1.0
1.0
1.0
0.9
0.9
0.9
0.8
0.8
0.8
0.7
0.7
0.7
0.6
0.6
0.6
0.5
0.5
0.5
88.0
96.0
104.0
1120
1200
87
90
89
92
33.6
36
2.8
2.8
2.3
2.3
2.0
2.0
1.7
1.7
1.5
1.5
1.3
1.3
1.1
1.1
1.0
1.0
0.9
0.9
0.8
0.8
0.7
0.7
0.6
0.6
0.5
0.5
112.0
120.0