11.0 CPR (Cardio Pulmonary Resuscitation)

To deflate mattress or overlay for CPR:
1. Disconnect the dual quick-connect mattress or overlay hoses from the Control Unit.
2. Begin CPR immediately. Do not wait for mattress or overlay to completely deflate.

12.0 System Cleaning, Care and Storage

12.1 Cleaning
Disconnect the AC power cord from the wall outlet before attempting to clean the Control
Unit. Do not heat or steam autoclave any component of the system.
Do not submerge the Control Unit.
1. To clean, use soap, water and a clean cloth to wipe down the Control Unit, power cord, hoses and
mattress or overlay. Do not use abrasive cleaners on the mattress. Wipe dry with a clean, dry cloth.
Note: Blood and other body fluids must be thoroughly cleaned from all surfaces before applying
disinfectants.
2. Gaymar surfaces are designed for reliability and long-lasting comfort. In order to maintain these
surfaces, they should be cleaned regularly using warm water and a hospital grade disinfectant. Gaymar
recommends Steris
3. Wipe down the mattress or overlay with a clean, dry cloth to remove any excess disinfectant.
4. Top Cover of mattress or overlay may also be laundered between patient uses or as required. Fill the
washing machine with warm water (70 -140°F or 21 - 60°C). Add one cup of laundry detergent. Place
no more than four top covers in a single extra large load capacity washing machine. When wash cycle
is complete, remove promptly from machine and ensure all excess water is drained from load. Place
dryer on LOWEST heat setting, or AIR FLUFF if available until dry. Verify top cover is completely dry
before placing under patient.
5. If individual air cells of mattress become soiled, clean and disinfect as described above or simply
replace air cell with a clean replacement. A plug is available from Gaymar to prevent water from
entering air cells during laundering (P/N 30287).
12.2 Storage
Control Unit
1. Check the AC power cord and plug for abrasions or excessive wear.
2. Coil power cord up and place both cord and Control Unit in plastic bag for storage.
Mattress/Overlay
1. Check the air manifold of the mattress/overlay for kinks or breaks. Replace if necessary.
2. To deflate, disconnect hose assembly from Control Unit.
3. Coil the hose assembly loosely and roll the deflated mattress or overlay starting at the head end.
When rolling up the mattress, use the straps for containment. Place in plastic bag for storage.
WARNING
®
®
Wescodyne
for both cleaning and disinfecting.