Herunterladen Inhalt Inhalt Diese Seite drucken

Safety Instructions - Grant XUBA Serie Bedienungsanleitung

Analogue ultrasonic bath
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 39

Safety instructions

Use of products
The following products are covered by this operating manual:
XUBA1, XUBA3
The products listed above are a series of general purpose ultrasonic cleaning baths
designed for indoor use by a professional user.
Electrical
Connect to a 230VAC fully earthed supply via a 3 pin plug. It can be dangerous to
operate an ultrasonic bath without an earth connected.
The mains plug is fitted with a 5A fuse. NEVER FIT A FUSE OF A HIGHER RATING.
Some baths are fitted with a detachable mains lead. These baths have a mains socket
on the rear of the bath fitted with two 20mm fuses. See rating information on page 15.
NEVER FIT A FUSE OF A HIGHER RATING.
Ensure that excess mains cable is stored neatly.
General use
Ensure that the bath is at least ¾ full of liquid before you switch it on.
Always use a basket to support items to be cleaned. Not doing so may damage the
bath and invalidate the warranty.
Do not place hands or fingers in the bath.
Care should be taken when operating the bath at higher temperatures as external sur-
faces may become hot.
Never use toxic, flammable, acidic, caustic or corrosive solutions in the bath.
Where the bath is fitted with a detachable mains lead disconnect the bath from the
mains supply before emptying. When emptying baths with no waste outlet, tip the bath
AWAY from the electrical connector and lead.
Read the whole of these instructions. Safety may be impaired if they are not followed.
Contact to Grant or its agent for advice on cleaning techniques and detergents.
Grant will not be responsible for damage or injury caused by incorrect use of the
equipment.
XUBA Series
2
29667 V1
Operating Manual

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Xuba1Xuba3

Inhaltsverzeichnis