Ultrasonic detergents

Detergents are a vital component in the ultrasonic process, aiding the removal and
loosening of debris from the surfaces of items placed in the tank while also intensify-
ing the power of the ultrasonic activity.
Grant offers a specially formulated ultrasonic detergent for use in many applications.
Grant M2 is a general purpose detergent for all plastics, glass and metals (expect
aluminium and other soft metals) which is also suitable for use on medical instru-
ments.
Dosing matrix (ml of detergent per tank)
Grant recommends a dosage of between 2 and 5% for all detergents.
XUBA1
XUBA3
Directions for use: Add 25-50ml of M2 per litre of water in the ultrasonic bath.
Warning: Grant M2 may cause etching to aluminium and other soft metals.
Shelf life: 6 months when stored in a cool, dark environment.
Precautions for use: In case of contact with eyes, rinse immediately with plenty of
water and seek medical advice. If swallowed, seek medical advice immediately and
show the bottle label. Keep out of reach of children.
Refer to safety data sheets available from Grant.
Grant M2 is biodegradeable
XUBA Series
Operating Manual
Detergent dose (ml)
2%
30
50
10
5%
75
125