Charging the Battery

IMPORTANT! Th e battery must be in an upright position during use and while charging.
IMPORTANT! Remove the charger from the carrying case before charging.
1.
Connect the auto adapter end of the charger into the battery pack socket, located in the front pouch of the carrying case.
See Figure 1.
2.
Plug the power cord into the charger.
3.
Monitor the LED on the charger while plugging the power cord into the wall outlet and ensure that it turns RED (the
charger LED may turn RED for only a short time before turning ORANGE). If the charger LED is RED for any length
of time, the charger is correctly connected to the battery. If the charger LED does not turn RED for any length of time,
leave the power cord disconnected from the wall outlet for 3 minutes and repeat this step. If the charger LED does not
turn RED, ensure that the auto adapter is fully seated in the battery pack socket and that the battery and charger fuses are
not blown (see Replacing the Battery Fuse and Replacing the Charger Fuse sections). If the charger LED still does not turn
RED, contact your home care provider. During normal operation, the battery charger's LED will appear RED while it is
charging. When the battery is 80-95% charged, the LED will appear ORANGE. When the charger is fully charged, the
LED will turn GREEN.
4.
After the LED turns GREEN, the charger can be disconnected from the battery.