6. Connection of the Spray Nozzle
1) Attach the Spray Nozzle (Optional) to the handpiece head
(Fig. 6).
2) Cut the Irrigation Tube to the desired length then connect
to the Spray Tube Nozzle on the motor (Fig. 6).
* Refer to "13. Option Parts List" to identify the correct Spray
Nozzle.
7. Checking the Handpiece Before Each Use
Follow the check procedure below before use. If any
abnormalities are found, stop using the handpiece
immediately and contact your Authorized B.A. International
Dealer.
1) Check the Head Cap is firmly tightened.
2) Insert the bur. (Refer to "5. Inserting and Removing the
Bur")
3) Rotate the handpiece for about one minute at the Max.
rotation speed of the attached motor. During rotation,
check for abnormalities such as abnormal rotation,
vibration, noise.
4) After the handpiece rotation has completely stopped, touch
the handpiece head to confirm the head is NOT heating
abnormally (Fig. 7).
9