Positioning the Freezpen
Hold the instrument vertically with the tip pointing
downward. Never use the Freezpen at an angle
of less than 45° from the vertical position.
Position the Freezpen correctly
Effective application
Before starting treatment consider the pathology,
location, size and depth of the lesion, the skin type,
the degree of tanning, the season of the year
and other parameters that may influence the result
of the treatment.
While using the Freezpen, the applicator tip with the
jet of cryogen should be held 0.5 to 3 mm from the
lesion. If the tip is held further than 5 mm from the
lesion, the liquefied N
O will expand into gas by the
2
time it reaches the tissue, which is not effective for
achieving therapeutic results.
By approaching the lesion with the jet already
flowing you can observe the focal point of the jet
on the tissue. The closer you get to the lesion,
the more freezing power you will apply to the tissue.
It is recommended to keep a greater distance for the
smaller and more delicate procedures.
The amount of liquid N
O that you wish to apply
2
can be controlled by moving the applicator quickly
towards and away from the tissue. Circular and
crossing movements have the same effect.
Do not hold the Freezpen applicator still over the
lesion. It may flood the area and damage healthy
11