Herunterladen Inhalt Inhalt Diese Seite drucken

Clinic6 freezpen Benutzerhandbuch Seite 10

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Installation of the assembled filter/opener device.
Step 3
I
Puncture the capsule
Keep the front end (tip of the Freezpen) pointing
downward. Gently screw the Freezpen back end to
the front end until you feel a little resistance. Now
do one last twist in a rapid, uninterrupted rotation
to puncture the capsule. You might hear a short
hissing sound. The shorter the hiss, the better!
Step 4
I
Use your Freezpen
Remove the applicator cap. Push the trigger to
start the flow of cryogen. Release the trigger to
stop the flow of cryogen. You may notice that
after releasing the trigger, the remaining gas in
10
the applicator is expelled. To switch the applicator,
unscrew the current one and screw the other one
on the Freezpen.
The table below gives the estimated application
with a continuous flow, time depending
on the applicator and the cartridge size.
The measurements were done at 21°C (70°F).
Freezpen 8g
Blue
White
Green
Freezpen 16g
110-140 s
250-280 s
80-100 s
180-210 s
145-165 s
Lesion size
1-3 mm
2-6 mm
4-8 mm