Lubrication (manual)
Inspection and function
2. Sterilization
(at the clinic validated proces-
ses as per ISO 17665-1:2006)
Storage
Aerosol spray
Insert the spray can nozzle into the motor input of the
handpiece and push the button of the can until the liquid
coming out from the head of the handpiece is clear. This ta-
kes a few seconds. Wipe off excess fluid with a paper towel.
Use only approved lubricants for dental handpieces.
Automated
Automatic reprocessing device for dental handpieces.
Visually inspect for appearance and function.
B-type steam
Pack in sterilization bag of paper/clear film (according to
applicable standards). The sterilization bag must be large
enough for the handpiece so that the bag is not stretched.
Pack individually and in dry condition. Check correct
sealing of the bag.
At least 3 minutes at
134 º C
Automated
Automated reprocessing device for dental handpieces.
Note: sterility is not preserved.
Chemklave
Not recommended.
Dry heat
Not recommended.
Protect from dust and dirt