Herunterladen Inhalt Inhalt Diese Seite drucken

Velocity Storage - phynix DC-2000C Bedienungsanleitung

Wanddickenmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

press
press
press
press
press
press
thickness value,
press
Smaller numerical values can be obtained by pressing this key several
times.
Repeat the last two steps until the displayed thickness value
corresponds to the actual thickness value d
Press
velocity.
press
press
press
This new velocity will be stored in one of the four locations. It can be selected
at menu 2. Velocity rate - 4. Vel.storage.
Alternatively, the speed of sound can also be calculated:
V
unknown

17.4.2.4. Velocity Storage

DC-2000 series provides 4 locations to store new velocities. Select one of these
as follows:
into 1. Measurement ,
or
to select 2. Velocity rate,
to confirm ,
or
to select 3. Vel. measurement,
to confirm ,
to move the cursor ▼ to the required digit position of the
to increase the numerical value at that position by 1.
to confirm. The display shows 4 locations to store this new
or
to select one location,
to confirm.
to get back to measurement.
= (v
x d
measurement
) / d
workpiece
display
.
workpiece
.
31

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für phynix DC-2000C

Inhaltsverzeichnis