Replace the cover and fit the appropriate location bosses.
The location boss for 1 litre vessels is supplied with the flask.
500 ml, 250 ml and 125ml flasks use the same location boss, as
supplied with the platform.

Siliconising Glassware

All glassware is siliconised before being dispatched. It will
eventually require resiliconising in order to maintain its water
repellant qualities.
CAUTION: THE WORKROOM MUST BE WELL VENTILATED.
Clean the glassware thoroughly and ensure it is free from grease.
Wash out and immerse the glassware in a water repelling
treatment such as Merck Ltd, "Repelcote VS".
Make sure the glassware is fully wetted by the solution.
Allow the solvent to evaporate. There will normally be sufficient
absorbed by water on the glassware to hydrolyse "Repelcote VS"
to a highly water repellent silicone film.