Fields located near RF
wireless communication
devices
Magnetic fields at
NOMINAL mains frequency.
) The device is destined for household use or use in professional healthcare facilities and may only be used in the patient's
a
room and in locations designated for respiratory therapy in a hospital or clinic. The most restrictive limits were applied
for the immunity tests.
b
) Before modulation.
) This test level presupposes a minimum distance of at least 15 cm between the piece of piece of ME EQUIPMENT or the
c
ME SYSTEM and the sources of electromagnetic fields at mains frequency.
Guidelines and Manufacturer's Declaration - Electromagnetic Immunity - Alternating Current Entry Port
Phenomenon
Electrical fast transient/
burst
Line-to-earth surges
Line-to-earth surges
Conducted disturbances
induced by radio
frequency fields.
Voltage dips.
Short interruptions
)
c
a
) RMS value before modulation.
) The frequency bands from 0.15 MHz to 80 MHz used by industrial, scientific and medical (ISM) radiofrequency devices are: 6.765
b
MHz - 6.795 MHz, 13.553 MHz - 13.567 MHz, 26.957 MHz - 27.283 MHz and 40.66 MHz - 40.70 MHz. The radio bands from 0.15 MHz
to 80 MHz used for amateur radio are: 1.8 MHz - 2.0 MHz, 3.5 MHz - 4.0 MHz, 5.3 MHz - 5.4 MHz, 7 MHz - 7.3 MHz, 10.1 MHz - 10.15
MHz, 14 MHz - 14.2 MHz, 18.07 MHz - 18.17 MHz, 21.0 MHz - 21.4 MHz, 24.89 MHz - 24.99 MHz, 28.0 MHz - 29.7 MHz and 50.0
MHz - 54.0 MHz.
) ME EQUIPMENT with direct current input for use with an AC/DC adapter, tested using an adapter compliant with the
c
specifications provided by the ME EQUIPMENT MANUFACTURER. THE IMMUNITY TEST LEVEL was applied at the adapter's
alternating current entry point.
IEC 61000-4-3
IEC 61000-4-8
EMC Regulation or test
method
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
COMPLIANT
NOTE: Further information about distances to be main-
tained between portable and mobile RF communica-
tions equipment (transmitters) and the Mister 8
02038211000000 can be requested from PIKDARE S.r.l. us-
ing the contact information provided in this manual. It is
nevertheless advisable to keep a safe distance (at least 1
m) between the electromechanical aerosol therapy device
and mobile phones or other RF communications devices in
order to minimise possible interference.
30 A/m
)
c
50 Hz or 60 Hz
Immunity test levels
Professional healthcare
facilities
± 2 kV
100 kHz frequency of repetition
± 0.5 kV, ±1 kV
Class II device or equipped with Class II AC/DC adapter.
Test not applicable
3 V
) 0.15 MHz - 80 MHz
a
6 V
) for ISM bands between 0.15 MHz and 80 MHz
a
80% AM at 1 kHz
0% in UT; 0.5 cycles
At 0°,45°,90°,135°,180°, 225°, 270° and 315°.
0% in UT; 1 cycle and 70% in UT; 25/30 cycles
Monophase: at 0°
0% in UT; 250/300 cycles
17
Household environment
)
b