Section 10: Troubleshooting
Although service and repair of the Cavitron BOBCAT Pro
Ultrasonic Scaler should be performed by DENTSPLY
personnel, the following are some basic troubleshooting
procedures that will help avoid unnecessary service calls.
Generally, check all lines and connections to and from the
ultrasonic scaler. A loose plug or connection will often create
problems. Check the settings on the ultrasonic scalerʼs knobs.

10.1 Troubleshooting Guide

:
System will not operate
1. Check that the ON/OFF Control switch is in the ON ( I )
position, and that the detachable Power Cord is fully
seated in the receptacle on back of the ultrasonic scaler.
2. Check that the ultrasonic scalerʼs three-prong plug
is fully seated in an appropriate AC receptacle, and
that AC current is present.
System operates:
(No water flow to insert tip)
1. Assure that water control is properly adjusted.
Section 11: Specifications
Cavitron ® BOBCAT ® Pro: 115 VAC Systems
Electrical Voltage
115V Gen-130
Gen-130 (Model approved for U.S. application)
Current
0.8 amps @ 115V (80VA)
Frequency
50/60 Hertz
Water Pressure
25–60 psig
Flow Rate
Minimum setting (CW) = 0 ml/min
Maximum setting (CCW) > 60 ml/min
Weight
6 lb. (2.7 Kg)
Dimensions
Height: 8 in.
Width: 3.8 in.
Depth: 8 in.
Handpiece Cable: 96 in.
Foot Control Cable: 96 in.
Power Cord: 96 in.
Water Supply Line: 96 in.
ENGLISH • 10
2.
Check that water supply control valve(s) (dental
office water supply) are open.
3.
If connected to a DualSelect Dispensing System,
check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
(Insert stops vibrating)
1.
Deactivate foot control.
2.
Verify insert is 25kHz and in good condition.
3.
Depress foot control.

10.2 Technical Support and Repairs

For technical support and assistance in the U.S. call
1-800-989-8826 or 717-767-8502 Monday through
Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time).
For other areas, contact your local DENTSPLY
representative.
Cavitron ® BOBCAT ® Pro: 220/230/240 VAC Systems
Electrical Voltage
Current
Frequency
Water Pressure
Flow Rate
Weight
Dimensions
220V Gen-130B
230V Gen-130U
240V Gen-130A, Gen-130E
0.4 amps @ 220V/230V/240V (80VA)
50/60 Hertz
172-414 kPa
Minimum setting (CW) = 0 ml/min
Maximum setting (CCW) > 60 ml/min
2.7 kg
Height: 20.3 cm
Width: 9.6 cm
Depth: 20.3 cm
Handpiece Cable: 2.4 m
Foot Control Cable: 2.4 m
Power Cord: 2.4 m
Water Supply Line: 2.4 m