Herunterladen Diese Seite drucken

Using The Microscope - Woodley Vision V5000 Bedienungsanleitung

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 34

4.0 USInG THe MICROSCOPe

4.1
Adjust the observation head
loosen the lock-screw, turn the observation head to a comfortable position for observation, and then
lock the lock-screw.
4.2
Place the specimen on the stage
lock the specimen slide on the mechanical stage using the slide clamp. Ensure that the specimen is
centred over the stage opening by adjusting the coaxial knobs of the stage.
4.3
Lamp settings
The microscope is fitted with power white lED illuminator. Before turning on the light, read the section
5.3 about electrical safety precautions. Insert the plug of the 7,5vdc power supply into the jack on the
rear and turn on the On/OFF switch. Turn the brightness adjustment knob to a brightness suitable for
observation.
4.4
Adjust interpupillary distance
hold the right and left parts of the observation head by both hands and adjust the interpupillary distan-
ce by turning the two parts until one circle of light can be seen.
4.5
Focus and diopter adjustment
rotate the coarse focusing knob to bring the slide into focus using an objective with a low magnifi-
cation. Adjust the fine focusing knob to get the image sharp and clear while observing with your right
eye, then turn the dioptric adjustment ring on the left eyepiece until a sharp and clear image can be
observed also with the other eye. Turn the tension-adjust-knob to get a suitable tension for the focus
system. When the image is in focus, turn the nosepiece to choose the objective you need.
4.6
Condenser adjustment
hold the notched ring at the base of the condenser and rotate it up or down until you reach a clear
and uniform sample illumination.
4.7
Set the numerical aperture
Adjust the aperture of the iris diaphragm to set
the numerical aperture of the illuminator, thus
controlling image contrast and resolution.
4.8
Additional filters
The green glass filter can be inserted in the flip-
out filter holder underneath the condenser.
4.9
Solar panel
If a solar panel is needed to work with microsco-
pe, please refer to the instructions enclosed with
solar panel kit. use the DC jack from back of mi-
croscope to plug the solar panel.
4.10
Fitting the mirror
If the microscope is supplied with a plano-conca-
ve mirror, this can be used with the microscope
instead of the electrical light, please follow these
instructions:
1) Make sure the equipment is not plugged into
the electrical power and the switch is in the OFF
position.
2) unscrew the light house from the illuminator,
(black part).
3) screw the mirror in place of the light house.
hOlD ThIs nOTChED rIng AnD
rOTATE IT
Page 7

Werbung

loading