Maintenance
Rules of use and identifying problems
■ When functionning, blu LED light and green or red LED show the battery level
■ If electrodes are dirty, damaged not well connected, the device doesn't work. Check the electrode
quality and connection, replace electrodes each 6 months
■ Control that electrodes are not contaminated or oxydated or not well connected.
■ If battery is low, recharge the unit.
Important :Maximum operating temperature is 50°C.
Green and red LED indicate the battery charge :
During the charge :
■ The device can't be used
■ Green LED flash
When charge is achieved :
■ Green LED are alight
Important : Never use liquid to clean the product. Protect it against all risks of liquid projection.
■
Battery
In order to preserve the durability of your battery, it is essential to respect the following regulations :
■ Never allow the battery to completely run out of charge.
■ Never store your device if the battery is uncharged.
■ In the case of prolonged storage without usage, recharge the battery at least every 3 months. The
battery will need to be charged more than this if the device is stored in areas greater than 25°C.
■ Store the device in an environment that is preferably between 0 and 20° Celsius
General functioning problems:
■ If the temperature of the battery is higher than the maximum value (lors de la charge ou en fonc-
tionnement): red LED light
►Leave the device to cool, if the problem persists, contact your distributor for a repair.
■ If any anomalies are detected on the circuit board: LED rouge s'alume
►contact your distributor for a repair.
■ If no LED are alight when trigger is pressed
►Charge the device
■ If the device still not working:
►contact your distributor for a repair.
Should you have any questions, please do not hesitate to contact us via email: mixplast@mix.fr